Note:
1. If use medical liquid of high viscosity and set low occlusion level, the machine may give occlusion
alarm when there is no occlusion in the tube, In this case, please observe the pressure indicator on the
screen and condition of syringe, and set higher occlusion level if necessary.
2. When high occlusion level is set, the high pressure inside the syringe may result in disconnection
of extension tube, please confirm extension tube is firmly connected to the syringe.
8.4.4 Key Lock setting
The optional setting including “Never”, “5min”, “4min”, “3min”, “2min” and “1min”.
★ Remark: Setting as “5min”/ “4min”/ “3min”/ “2min” or “1min” means all keys shall be locked (except
POWER key) after“5min”/ “4min”/ “3min”/ “2min” or “1min” if no operation on keys.
To unlock the keypad: press OK + clear key together.
8.4.5 Bolus setting
There are two methods for Bolus, manual Bolus and auto Bolus, operation as below:
Manual Bolus: After power on the pump, press STOP+ left 1st soft key to enter Settings, select
Bolus, set Bolus rate and VTBI: 0, press ENTER to save the value. During infusing, keep pressing
BOLUS key then it will run at Bolus rate; if release the BOLUS key, it will run at original rate, not Bolus
rate.
Auto Bolus: After power on the pump, press STOP+ left 1st soft key to enter Settings, select Bolus,
set Bolus rate and VTBI >0, press ENTER to save the value. During infusing, press BOLUS and release,
it will run at Bolus rate until finishing the VTBI, then run at original rate.
Press to turn to next page.
8.4.6 Back light setting
There are two methods for Bolus, manual Bolus and auto Bolus, operation as below:
Manual Bolus: After power on the pump, press STOP+ left 1st soft key to enter Settings, select Bolus,
set Bolus rate and VTBI: 0, press ENTER to save the value. During infusing, keep pressing BOLUS key
then it will run at Bolus rate; if release the BOLUS key, it will run at original rate, not Bolus rate.
Auto Bolus: After power on the pump, press STOP+ left 1st soft key to enter Settings, select Bolus,
set Bolus rate and VTBI >0, press ENTER to save the value. During infusing, press BOLUS and release, it
will run at Bolus rate until finishing the VTBI, then run at original rate.