syringe pump serial number and then click “start”.
e. Press to upload records to computer terminal; press “ transfer- capture text - Stop” after
finishing uploading, then all infusion/alarms records can be reviewed on the txt that established previously.
After finishing uploading, the syringe pump will return to the previous menu automatically.
(2) View Log: Press or to view the latest infusion/alarm information. Press to check
previous records or for next records. The alarm will not be recorded that it is closed by the system.
Press ‘Back’ to return to the previous menu.
(3) When the pump is power off, the history records are still kept inside. After shut down, the
electronic memory is 5years.
8.4.11 Date and Time setting
Press OK to enter ‘Date and Time’ setting interface. The editable value is in white shade. Press
or to input parameters and press OK to save. Set the value in year, month, day, minute and
second digits in sequence with this method.
Press to turn to next page. Press to turn to last page.
8.4.12 Auto shutdown setting
Select “never” or “1-99” minutes according to the requirements and press OK key to save the
parameters "1-99" minute means that in the non-infusion non-alarm state, Infusion pump automatically
shutdown after "1-99" minutes without operation.
8.4.13 Night mode setting
Select “ON”(turn on night mode) or “OFF”(turn off night mode).
If setting as ON, shall display on LCD. Key sound shall be off; the screen shall turn dark after
1 minute if no operation on keys; the top indicator light shall be off during infusion. (if there is any alarm,
the indicator light shall be on.)
8.4.14 Language setting
After entering “Language” setting interface, press or to select required language, then press
OK to save setting.
8.4.15 Bed number setting