Electrical connections
Transducer connection
30 A03470_07_E00_00 HBM: public DMP41
4.2.3 Temperature sensors
Cable wire assignment when using an RJ45 connector:
Cable color code: wh= white; bu= blue;
Connection diagram
Pin assignment
Connection point
PIN 1, 3, 5, 7
PIN 2, 4, 6, 8
1
8
GND
DATA
bu
wh
RJ45 connector
4 x
SYNC. IN
SYNC. OUT
DIG I/O
RJ45
PIN Designation
1
2
3
4
5
6
7
8
GND
Temp 1
GND
Temp 2
GND
Temp 3
GND
Temp 4
WARNING
An electric shock can be fatal or cause serious injury
Make sure that all the metallic surfaces to which the temperature sensor
is attached (glued, clamped) are voltage-free.