WE2111
18 A3983-1.0 en/de/fr
Connector socket for load cell(s)
Fig. 6.1 View from rear, connection of load cell(s)
"
Connect a 9-pin D-Sub connector as shown in the
table with the connections of the load cell(s).
Use shielded cables for the connecting cables and
connect the shield extensively to the connector
housing.
Important
If the load cell(s) are in four-wire configuration, the sense
leads must be connected to the excitation voltage with
jumpers: 2 (+) to 1 (+) and 4 (-) to 3 (-).
The sense lead inputs must not remain open otherwise
only error messages will be displayed (E00040, E00080
or E000C0).