WE2111
32 A3983-1.0 en/de/fr
version (e.g. P60c) and calibration counter (e.g.
C00008).
10.4 Calibration
The display must show buiLd before the following
settings can be made.
"
Press the button twice.
CAL is displayed.
"
Press the button.
ZErO is displayed.
"
Press the button.
The current measured value is displayed.
"
Relieve the scale and press the button.
Z in P is displayed as long as the zero balance is
being performed. 0.00 flashes in the display when the
zero balance has been performed.
"
Press the button.
ZErO is displayed. The zero balance is then
complete.
"
Press the button.
SPAN is displayed.
"
Press the button.
The current measured value is displayed. The display
flashes for you to enter the weight used.