WE2111
A3983-1.0 en/de/fr 33
"
Use the and buttons to enter the weight used,
10 kg in this case.
"
Load the scale with 10 kg and press the button.
S in P is displayed as long as the zero balance is
being performed. 10.00 flashes in the display when
the adjustment has been performed.
"
Press the button to quit the menu.
"
Press the button for several seconds to save the
settings.
The following are displayed in sequence before the
actual measured value appears: SAviNG, firmware
version (e.g. P60c) and calibration counter (e.g.
C00008).