9/31
5 Error messages
The error message will be indicated in the speed display of the front panel
e.g. :
RPM x 100
t
START
STOP
5.1 MAINS RESET
Switch off the mains switch.
Wait at least for 10 seconds and then switch on the mains switch again.
5.2 Brief description
Display. Fault Brief description Page
- 1 - Tacho error Tacho pulses break down 10
- 2 - System reset Mains interrupt 10
- 3 - Imbalance Imbalance on the motor axis 10
- 4 - Communication Communication error 11
- 5 - Overload Fault in the motor or the motor control 11
- 7 - Overspeed Fault in the power board 11
Versions Error Wrong machine version setting 12
- d - Lid lock error Fault in lid lock system 12