Hettich ROTOFIX 32 - Page 13

Hettich ROTOFIX 32
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13/34
– 8 – Undervoltage
Error Mains voltage less than 20% as nominal voltage.
Error
consequence
Drive switches off. No brake effective
Error cause
Mains voltage too less
Power board (A1) defective
Measurement Check the mains voltage
Measure intermediate voltage, see chapter 9.2.6
Error code reset Perform a MAINS RESET after standstill.
– 9 – Overtemperature
Error Overtemperature in the motor indicated.
Error
consequence
Drive switches off. No brake effective
Error cause
Motor defective
Power board (A1) defective
Overtemperature switch in motor defective
Loosen contact in plug S401 Pin 4 or 5
Measurement
Remove plug S401 and measure between Pin 4 and 5 :
Switch closed: 0 OK
Switch open: defective
See also chapter 9.2.
Error code reset Perform a MAINS RESET after standstill.