– c – Controller-Watchdog
Error Watchdog in power board
Discrepancy in program procedure
Error
consequence
Drive switches off. No brake effective
Error cause
Power board (A1) defective
Error code reset Perform a MAINS RESET after standstill.
– d – Lid lock error
Error Lid lock is open during centrifugation.
Error
consequence
Drive switches off. No brake effective
Error cause
Micro switch on lid lock is defective.
Power board (A1) defective.
Loosen contact in Plug S404 Pin 5 and 6.
Mechanical defect at the lid lock
Emergency release active during run.
Measurement Remove plug S404 and measure between Pin 5 and 6 :
Lid closed: 0
Lid open:
See also section 9.2.
Error code reset Perform a MAINS RESET after standstill.
– E – Short circuit
Error Power board detects short circuit
Current consumption too high
Error
consequence
Drive switches off. No brake effective
Error cause
Power board (A1) defective
Error code reset Perform a MAINS RESET after standstill.