EasyManua.ls Logo

Hill-Rom Synclara PSC1 - Page 52

Hill-Rom Synclara PSC1
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Device Settings
NOTE:
The pressure management settings show:
• Pressure Ceiling (PC)—the maximum therapy pressure that can
be selected by the user in manual mode or when modifying a
preset plan.
• Highest Programed Pressure (HPP)—the highest therapy
pressure that has been programed in both automatic and
manual modes in this system.
2. Press the + or - controls to select the required PC setting.
NOTES:
• Each press of the control increases/decreases the PC setting by
five (5) units.
• The PC ranges between 5 - 70 cmH2O and is limited to the HPP
gathered from the system.
• The PC must be equal or greater than the HPP.
• For example,
If HPP equals 45 cmH2O, then PC must be equal to or
greater than 45 cmH2O. The user can not select a PC setting
of 44 cmH2O or lower.
3. When complete, slide the Clinical
Access setting to OFF. This prevents
unauthorized persons from adjusting the
settings.
Draft_E_2019-Apr-29_Cleaned

Table of Contents