4.1 Comparator Function
56
2
Display the absolute value judgment function configuration screen.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16)
3
Set the absolute value judgment function to either "On" or "Off."
on
.........................
Absolute value judgment function on.
oFF................... Absolute value judgment function off.
4
Accept the setting and return to the measurement screen.
For example, the following judgment results would be obtained when connecting
the probes backwards to a 3.7 V battery (resulting in a displayed voltage mea-
sured value of -3.7 V) with an upper threshold of 3.9 V and a lower threshold of
3.6 V:
Absolute value judgment function off: Lo
Absolute value judgment function on: IN
Enabling and Disabling the Comparator Function
Enables the comparator
When the comparator is enabled, the following key operations are disabled
to avoid inadvertent operations.
V//V key (Measurement mode setting)
SHIFT
V//V key (Zero-Adjustment)
SHIFT COMP key (Comparator setting)
AUTO key (Auto-ranging setting)
SMPL key (Sampling rate setting)
SHIFT
SMPL key (Averaging setting)
SHIFT TRIG key (Trigger source setting)
SHIFT
ENTER key (Menu display)
SHIFT
STAT key (Delay setting)
Range keys
COMP lit
When the comparator is enabled, auto-ranging is automatically disabled.