4.11 Measurement Value Output Function
74
This function causes output of measured values via the RS-232C or LAN inter-
face in the same sequence as trigger input.
This function is useful when measuring using internal (free-run) triggering, and
for obtaining measured values on a PC when using a footswitch for triggering.
4.11 Measurement Value Output Function
1
(The SHIFT indicator lights up.)
The Menu display appears.
2
The Measurement Value Output function setting display appears.
See "1.4 Menu Display Sequence (SHIFT ENTER)" (p.16).
3
Turn Measurement Value Output Function on or off.
on.......... enables the measurement value output function on.
oFF........ disables the measurement value output function off.
4
Applies setting and returns to the Measurement display.
5
The measured value is output from the RS-232C or LAN interface
when you press the
TRIG key or when a signal is applied to the EXT
I/O TRIG
terminal.
Set the PC to the receiving state beforehand. When a measurement
value is received, the PC should perform appropriate processing
such as recording or displaying, then re-enable the receiving state.
(Main Display)
(Sub Display)
The current setting blinks.
When external triggering is enabled, a measurement is performed and the
value is sent after each trigger event. When internal triggering is enabled,
the first value measured after triggering is sent.
The measurement output function is not applicable to the GP-IB interface or
printer.