Bump Test via SSDC
Apply gas from a cylinder to the sensors manually through the calibration cap, and using the
Safety Suite Device Configurator (SSDC) software on a personal computer (PC).
1. Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs to
snap it into place. Wait a few minutes to sensors warm up.
2. Attach the hose.
3. Connect the instrument to the PC via the IR link.
BW Icon 21 User Manual