Time Out Events
Action Time Out
Auto exit error screen and power off 5 seconds
Auto skip error message screen and enter warmup 5 seconds
Auto exit menu and turn off Icon LED 6 seconds
Auto exit Force bump and Calibration 30 seconds
Exit auto detected span gas 60 seconds
Pairing, bump, and calibration result display 5 seconds
BLE pairing timeout 60 seconds
Troubleshooting
Problem Cause Solution
Battery icon blinks for 5
seconds when press
button to power on.
Depleted battery
Charge the rechargeable
battery pack
The detector, side LEDs, all
bays, and IntelliFlash blink
for 5 seconds when press
the button to power on.
The detector expired
The apparatus is over two
years lifetime, cannot
continue to use.
All bays and IntelliFlash
light for 5 seconds
All sensors fail
Replace the sensor or the
PCBA
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound two
long beeps.
RTC fail Replace PCBA
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound five
short beeps.
Data flash fail Replace PCBA
The detector, side LEDs,
and IntelliFlash light for 5
seconds, and sound one
long beep and two short
Temperature sensor fail Replace PCBA