Scale Calibration Procedure (lbs. or kg) with Remote Display
7. Make sure there is no load on the scale platter.
8. Calibrate a zero load. Press the right arrow () button once when the remote display reads CAL 1 and
there is no load on the scale platter.
Figure 42.
9. Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter then press the right arrow ()
button.
Figure 43.
10. Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to simulate a full load, center entire
load then press the right arrow () button.
Figure 44.
11. Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load then press the right arrow ()
button.
Figure 45.
48