Scale Calibration Procedure (lbs. or kg) with Remote Display
12. Remove the remaining half load from the scale then press clear (C). The message done will flash briefly
on the display.
Figure 46.
13. Exit calibration mode. Press and hold the test button for at least 3 seconds then release. With no load on
the scale the display should read 0.00 lb. or 0.000 kg.
Figure 47.
14. The unit’s calibration must now be verified as required by state and/or local Weights and Measures
regulations (starting on page 54).
Need to Start Over?
To exit the calibration mode or restart the process press the test button then the clear button.
Figure 48.
49