Manual” carefully to ensure that the minimum voltage of the detector is not less than 16VDC.
5.4) Grounding connection
In order to ensure the electromagnetic compatibility and anti-interference ability of the
instrument, good grounding is essential. For detailed use and connection, please refer to the
User's Manual.
The shielding layer of cable must follow the principle of "single point grounding" Usually, the
shielding layer of the cable at the transmitter end is connected to the common point of the
instrument, that is, the shielding layer wire is connected to an empty terminal outside or
inside the instrument.
6) System Status
RAEGuard Lite provides a clear, definite and user-friendly human-machine interface with
three magnetic keys, one main LCD screen and three system status LED indicators. For
user's convenience and quick use of the instrument, the following table lists the system
status LED light indicators and corresponding current output values. For more information,
see "User's Manual".
Note: Fault, inhibit, and warning current set to: 1mA, 2mA, 3mA; overrange current set to 22 mA. And can’t be modified.