I
-
I
-
Error Messages
SCPI Error Messages
-170
“Expression error”
This error, as welI as errors -171 through -179, are generated
when parsing an expression data element. This particular error
message should be used if the device cannot detect a more specific
error.
-171
“Invalid expression”
The expression data element was invalid. (See IEEE 488.2,
7.7.7.2); for example, unmatched parentheses or an illegal
character.
-178
-
180
“Expression data not allowed”
A legal expression data was encountered but was not allowed by
the device at this point in parsing.
“Macro error”
This error, as well as errors -181 through -189, are generated
when
dehning
a macro or executing a macro. This particular error
message should be used if the device cannot detect a more specific
error.
-181
“Invalid outside macro definition”
Indicates that a macro parameter placeholder
(knumber)
was
encountered outside of a macro dehnition.
-183
-184
“Invalid inside macro definition”
Indicates that the program message unit sequence, sent with a
*DDT or *DMC command, is syntactically invalid. (See IEEE 488.2,
10.7.6.3).
“Macro parameter error”
Indicates that a command inside the macro
dehnition
had the
wrong number or type of parameters.
16-12