Error Messages
SCPI Error Messages
421
-330, “Self-test failed;Power Up Calibration Defaulted(421)”
Indicates that after the instrument is up and running, a series
of power on self-tests have been run and error correction code
checking has found that contents of one of the calibration arrays
were found corrupted. A default calibration has been used.
422
-330, “Self-test failed;Power Up Calibration Improved(422)”
Indicates that after the instrument is up and running, a series
of power on self-tests have been run and error correction code
checking has corrected a one bit error when recovering date from
the EEPROM. Proper operation of the instrument is guaranteed.
It is suggested that a calibration save operation be performed to
permanently correct this problem.
423
-330, “Self-test failed;Power Up DSP Handshake Failed(423)”
Indicates that during the series of power on self-tests, the
handshake control with the Digital Signal Processor has failed.
424
-330,“Self-test
failed;DSP Handshake Fail During Byte
Transfer(424)”
Indicates that after the instrument is up and running, byte
transfers with the Digital Signal Processor have failed.
430
-
120, “Numeric data error;Entered Value is not a Valid Patch
Number(430)’
Indicates that a Self-Test patch had been requested to be deleted
from the eeprom Self-Test Patch
‘lhble,
however, the entered value
is not a valid patch number. A valid patch number is the index of
the patch item in the patch table (starting at 1.) It is not the test
<name>.