I
-
I
-
Error Messages
SCPI Error Messages
414
415
416
417
418
419
420
Indicates that after the instrument is up and running, a series
of power on self-tests have been run and error correction code
checking has found that FLASH ROM has a high byte error.
-330, “Self-test failed;Boot-ROM Checksum Error (LOW
BYTE)(414)”
Iudicates that after the instrument is up and running, a series
of power on self-tests have been run and error correction code
checking has found that Boot-ROM has a low byte error.
-330, “Self-test failed;Boot-ROM Checksum Error (HIGH
BYTE)(415)”
Indicates that after the instrument is up and running, a series of
power on self-tests have been run and error correction haming
code checking has found that Boot-ROM has a high byte error.
-330, “Self-test failed;RAM-backup battery is
LOW(416)”
Indicates SRAM-backup battery is LOW.
-330, “Self-test failed;Power Up RAM Addressing
Error(417)”
Indicates RAM Addressing Error during Power Up.
-330, “Self-test failed;Power Up RAM Test Error(LOW
BYTE)(418)”
Indicates that after the instrument is up and running, a series of
power on self-tests have been run and RAM Test is found to have
low byte error.
-330, “Self-test fa.iled;Power Up RAM Test Error(HIGH
BYTE)(419)”
Indicates that after the instrument is up and running, a series of
power on self-tests have been run and RAM Test is found to have
high byte error.
-330, “Self-test
failed;Power
Up Calibration Corrupted: Default
Used(420)”
Indicates that after the instrument is up and running, a series
of power on self-tests have been run and error correction code
checking has found that contents of one of the calibration arrays
were found corrupted. A default calibration has been used.
16-30