ColorFlex EZ User’s Manual Version 2.2
10
Figure 3. Main Menu
When OFF is shown as an available button command on the main menu screen, choose OFF to
immediately turn the instrument off.
About the Instrument
Information specific to the instrument may be viewed on the ABOUT YOUR INSTRUMENT
screen, which can be obtained by choosing ABOUT from the main menu. The instrument
model, serial number, firmware version, and total flash count for this instrument lamp are
displayed.
Figure 4. About Your Instrument
More details on ColorFlex EZ operation are given in Chapters 2 through 4.
Taking a Simple Measurement
This chapter describes the various steps performed in the course of normal instrument
operations, such as standardizing, reading, storing and printing data.
Standardization sets the top and bottom-of-scale for the neutral axis. During standardization,
the bottom-of-scale (zero) is set first. This is done by placing the Black Glass at the sample port.
The top-of-scale is then set by reading the calibrated White Tile. Messages on the display
prompt through the standardization process as described below. The time elapsed (in minutes)
since the last standardization is shown in the upper right corner of the screen.