Checking the System
You may check your meter and test strips using the
TEE2 Control Solution(control A and/or B).
The TEE2 Control Solution contains a known amount
of glucose and is used to check that the meter and the
test strips are working properly.
The test strip vials have TEE2 Control Solution ranges
printed on their labels. Compare the result displayed
on the meter to the TEE2 Control Solution range
printed on the test strip vial.
Before using a new meter or a new vial of test strips,
you may conduct a control solution test following the procedure
on pages 19-20.
Notes:
• Useonlythe
TEE2
Control Solution (available for purchase
separately).
• Checktheexpirationdatesprintedonthebottle.Whenyoufirst
open a control solution bottle, record the discard date (date
opened plus three (3) months) in the space provided on the label.
• Makesureyourmeter,teststrips,andcontrolsolutionareat
room temperature before testing. Control solution tests must be
done at room temperature (20-25˚C, 68-77˚F).
• Beforeusingthecontrolsolution,shakethebottle,discardthe
first few drops and wipe the tip clean.
• Closethecontrolsolutionbottletightlyandstoreata
temperature of 8-30˚C(46-86˚F).
Manufacturer : i-SENS, Inc.
27-36, Gwangun-ro, Nowon-gu, S
Discard Date :
L O T
PGA2
Glucose Contro
TM
You may do a control solution test:
• Whenyouwanttopracticethetestprocedureusingthecontrol
solution instead of blood,
• Whenusingthemeterforthefirsttime,
• Wheneveryouopenanewvialofteststrips,
• Ifthemeterorteststripsdonotfunctionproperly,
• Ifyoursymptomsareinconsistentwiththebloodglucosetest
Control Solution Testing
Step 2
Shake the TEE2 Control Solution
bottle before each test.
Remove the cap and squeeze the
bottle to discard the first drop. Then
wipe the tip with a clean tissue or
cloth. After the
symbol appears
on the display, apply the solution to
the narrow edge of the test strip until the meter beeps. Make
sure the confirmation window fills completely.
Step 1
Insertateststripintothemeter’steststripport,
with the contact bars facing upwards.
Gently push the test strip into the port until the
meter beeps. Be careful not to bend the strip
while pushing it in.
The
symbol will be displayed on the screen.
Note:
•The meter may switch off, if the control solution sample is
not applied within 2 minutes of the
symbol appearing on
the screen. If the meter turns off, remove the strip, reinsert,
and start from step 1.
results and you feel that the meter or test strips are not working
properly,
• Ifyoudropordamagethemeter.

Questions and Answers: