Note:
•The TEE2 Control Solution can be purchased separately.
Please contact Spirit Healthcare on 0800 881 5423.
Caution:
•The range printed on the test strip vial is for the TEE2
Control Solution only. It does not have any connection to
your blood glucose level.
Step 3
A test result will appear after the meter counts
down from 5 to 1.
After your control solution result appears on
the display, press C
for3secondstillthe‘check’
symbol appears on the display.
Whenthe‘check’symbolisdisplayed,theresult
isstoredinthemeter’smemoryandisnot
included.
Step 4
Compare the result displayed on the meter
to the range printed on the test strip vial.
The result should fall within that range.
Used strips should be discarded safely in
disposable containers.
Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur due to the following factors:
If results continue to fall outside the range printed on the test
strip vial, the TEE2 Test Strip and Meter may not be working
properly. Do not use your system and contact Spirit Healthcare
24/7 Freephone Support Line at 0800 881 5423.
Situations
•Whenthecontrolsolutionbottlewas
not shaken well,
•Whenthemeter,teststrip,orthecontrol
solutionwereexposedtohighorlow
temperatures,
•Whenthefirstdropofthecontrol
solution was not discarded or the tip of
the bottle was not wiped clean,
•Whenthemeterisnotfunctioning
properly.
•Whenthecontrolsolutionispastthe
expirationdateprintedonthebottle,
•Whenthecontrolsolutionispastits
discard date (the date the bottle was
opened plus three (3) months),
•Whenthecontrolsolutionis
contaminated.
Actions
Repeat the
control solution
test by referring
to the “Notes”
on page 18.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.