30
ACCURA9/9L INSTRUCTION MANUAL
5
D.Sam
FunctionSettin
Screen
***D.SampMenu***
Ch00BotLevelON
TopLevelNOD
EmpLevelOFF
Sampling at the top level:
Check if Bot-Level is ‘ON’ at D.SAMP setting
function screen.
Put the cursor on ‘Top-Level’.
Press
CTRL+RETURN key
.
The next screen will appear.
6
Sam
lin
Screen
**
**
*DATASAMPLING!!*
********************
Set the barrel filled with material up to the upper limit
and press
RETURN key
.
Press
CLR RESET
key to stop.
7
ƒgƒbƒv Samp‰
TopSamplingscreen
***TopLevel***
Ch00
Vc.Press015.0
Vt00.500Vi01.000
Enter the vacuum pressure (Vc.Press), time (Vt.) and
interval time (Vi) without material dropping and
bubbling.
Input the Vt and Vi more than 0.030 sec.
If O.K., press the
RETURN key
.
Press
SHIT+CLR RESET key
to stop.
8
SamplingScreen
***SamplingMode**
*
*DATASAMPLING!!*
********************
The dispensing for the downloading of sampling data
is carried on a few times.
During the sampling, this screen is shown.
After the sampling, D.SAMP screen is shown.
The procedure at the upper level of barrel is
completed.
9
D.SampFunctionSettingScreen
***D.SampMenu***
Ch00BotLevelON
TopLevelON
EmpLevelOFF
Check if the Bot_Level and Top_Level are ‘ON’ at
the D.SAMP function setting screen.
Put the cursor on the Emp_Level ‘OFF’ by setting
the cursor key
(↑、↓)
unless using the empty
alarm function.
Press
CTRL+RETURN key
while putting the cursor
key on Emp_Level if using the empty alarm function.
The next screen will appear.