EasyManua.ls Logo

IFM Efector500 PY7102 - Parameter Setting; 8�1 Parameter Setting General

IFM Efector500 PY7102
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
UK
8 Parameter setting
During the parameter setting process the unit remains in the operating mode� It
continues its monitoring function with the existing parameters until parameter set-
ting has been terminated�
8.1 Parameter setting general
Each parameter setting requires 3 steps:
1
Selecting parameter
Press [Mode/Enter] until the re-
quested parameter is displayed�
 
2
Setting the parameter value
Press[Set]andkeepthebuton
pressed�
Currentsettingvalueoftheparam- >
eter bit flashes for 5 s�
After 5 s: Setting value is changed: >
incremental by pressing briefly or
scrolling by holding pressed�
 
The numerical values are incremented continuously� If the value is to be reduced: Let
the display move to the maximum setting value� Then the cycle starts again at the
minimum setting value�
3
Acknowledge parameter value
Press [Mode/Enter] briefly
The parameter is displayed again� >
The new setting value is stored�
 
Set more parameters:
Start again with step 1�
Finishing parameter setting:
Press [Mode/Enter] several times until the current measured value is displayed
or wait for 15 s� The unit returns to the operating mode if no button is pressed for
over15safteracknowledgementofthenewparametervalue.

Related product manuals