OPERATOR'S MANUAL
25
6.1. SETTING THE OPERATING MODE
X-ray unit features the following operating modes:
- AUTO: the system automatically selects the best setting available for each tooth-patient size combination
- USER: the system automatically proposes the optimum exposure time according to the selected tooth and patient
size - En65: all exposures are performed at 65KV and 6mA.
For both operating modes, the exposure time is in the range 0.02s - 1s. The permitted anode voltage and current
combinations are shown in the following table:
NOTE: This section applies only to models RX DC REF: I3PV*6***
(Character * can be any alphanumeric value)
Table 1
60-7
60 kV
7 mA
63-6
63 kV
6 mA
65-6
65 kV
6 mA
NOTE: This section applies only to models RX DC REF: I3PV*7***
(Character * can be any alphanumeric value)
Table 2
60-8
60 kV
8 mA
65-8
65 kV
8 mA
70-8
70 kV
8 mA
60-4
60 kV
4 mA
65-4
65 kV
4 mA
70-4
70 kV
4 mA
To set the operative mode, press keys and to go to the setup menu.
Scroll the parameters up to parameter P06 and press key . Scroll the options to find the desired operating
mode and then press key .
Press key to quit the setup menu.