X-RAY KEY RELEASED TOO EARLY
Hold down the key until the image has been
captured.
SHOOTING SEQUENCE NOT COMPLETED
Handheld most likely lost the signal. Try to repeat
exposure. If the problem persists, contact the
technical service department.
HANDHELD INTERNAL TEST ERROR
Take out the batteries and then put them back in
after waiting a few seconds. If the problem
persists, contact the technical service department.
HANDHELD AUTO DIAGNOSIS TEST
FAILED
Contact technical service department.
Try to repeat exposure. If the problem persists,
contact the technical service department.
Handheld lost the signal. Try to repeat exposure.
If the problem persists, contact the technical
service department.
HANDHELD SERIAL NUMBER INCORRECT
OR NOT INITIALIZED
Contact technical service department.
X-RAY UNIT INTERNAL ERROR
Contact technical service department
COLLIMATOR SELECTION NOT
CONSISTENT
After turning the rectangular collimator on or off,
wait a few seconds to allow the icon on the
handheld to be updated.
Contact technical service department
Wait approximately 15 minutes for automatic
system reset
Check the supply system. If the problem persists,
contact the technical assistance department.
SUPPLY VOLTAGE TOO HIGH/LOW
Repeat the x-ray. If the problem persists, contact
the technical assistance department.
INTERNAL ADJUSTMENT PROBLEM
Reduce the distance between the remote control
and x-ray head and then repeat the x-ray. Follow
the information given on how to properly use the
handheld’s antenna. If the problem persists,
contact the technical service department.
X-ray generator or arm cord may be faulty.
Contact technical service department.
Close the door and if necessary repeat the test.
NOTE: Just press button to exit error condition
NOTE: As regards the other error codes, CONTACT the technical service department.