OPERATOR'S MANUAL
EN
8. PERIODIC MAINTENANCE
WARNING:
Any technical maintenance work required must be carried out by qualified personnel or by a specialised
technician authorised by the manufacturer. It is the user’s responsibility to check that an authorised
technician carries out routine maintenance at least every year. The maintenance methods are specified in
the Technical Service Manual possessed by the Authorised Technicians
For safety reasons and for the health of the patient, operator and third-parts inspections and maintenance must be
carried out at regular intervals.
Period
Operator
Object
Description
1 year
Specialised technical
distributor who installed the
device or other technicians
authorised by the
manufacturer
All device
components are
integral parts of the
unit
In order to ensure the safety of
operation of the device, it is advisable
to inspect the x-ray unit in all its parts,
in order to prevent or repair any faults
Quality control by means of a dental phantom for image acquiring systems, according to IEC 61223-3-4:2000:
Image resolution (lp/mm)
4lp/mm
Low contrast resolution
2
Artifacts
In the image, there must be no artifacts such as visible
horizontal lines
Control period
1 year
The quality control consists in performing a radiological investigation by means of the tested device and the
acquisition system Zen-X or X-Pod. The sensitivity of the x-ray unit must be set according to what is stated in
paragraph 14.
Alternatively, you can verify that the measured load factors (kV, mA, ms) fall within the accuracy limits stated in
Section 11.
Periodic monitoring ensures the proper functioning of the device and the conformity of the results obtained.