Transmit: DCB [,a,b] <CR>[<LF>]
Description
a Measurement channel, a =
0 –> Measurement channel 1
1 –> Measurement channel 2
2 –> Measurement channel 3
b Bar graph display, b =
0 –> Off (default)
1 –> Bar graph covering full scale range
2 –> Bar graph covering full scale range, high-level
presentation
3 –> Bar graph covering full scale range and setpoint
threshold
4 –> Bar graph covering a decade according to current
measurement value
5 –> Bar graph covering a decade according to current
measurement value, high-level presentation
6 –> Bar graph covering a decade according to current
measurement value and setpoint threshold
7 –> p = f
(t)
, autoscaled, 0.2 seconds / pixel
For each measurement every 200 ms a measurement
value is saved in tabular form and the last 100
measurement values (=100 pixel) are shown
autoscaled.
The represented data string corresponds to a logging
duration of 20 seconds.
8 –> p = f
(t)
, autoscaled, 1 second / pixel
For each measurement every second a measurement
value is saved in tabular form and the last 100
measurement values (=100 pixel) are shown
autoscaled.
The represented data string corresponds to a logging
duration of 100 seconds.
9 –> p = f
(t)
, autoscaled, 6 seconds / pixel
For each measurement every 6 seconds a
measurement value is saved in tabular form and the
last 100 measurement values (=100 pixel) are shown
autoscaled.
The represented data string corresponds to a logging
duration of 10 minutes.
10 –> p = f
(t)
, autoscaled, 1 minute / pixel
For each measurement every minute a measurement
value is saved in tabular form and the last 100
measurement values (=100 pixel) are shown
autoscaled.
The represented data string corresponds to a logging
duration of 100 minutes.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b <CR><LF>
Description
a Measurement channel
b Bar graph display
DCB - Display control bar graph