INFICON Installation | 5
XL3000flex-Operating-instructions-jina83en1-07-(2105) 31 / 108
ü You have a separately available I/O module or bus module. See also "Accessories
[}101]".
ü You have a commercially available DIN-TS35 DIN rail.
1
Place the sniffer leak detector at least 20 cm away on all sides.
2
Connect the SL3000XL sniffer cable to the front of the XL3000flex.
3
Fasten the DIN rail using the screws already located in the threaded holes on
the back of the device (see Device setup [}14]).
4
Then hook the module into the DIN rail at the bottom, press it against the DIN
rail at the top and let it snap into place.
5
Connect the I/O module or the bus module via a data cable to the LD socket on
the back of the XL3000flex.
See also
2 Selecting the type of expansion module [}45]
2 Assigning analog outputs of the I/O module [}46]
2 Settings for bus module BM1000 [}53]
5.4 Fasten holder for sniffer line (optional)
WARNING
Danger to pacemaker wearers from magnets
The function of a pacemaker can be influenced by the magnet on the back of the
holder.
► If you wear a pacemaker, do not install it yourself.
► If you wear a pacemaker, always maintain a distance of at least 10 cm from the
holder when operating the device.
CAUTION
Risk of injury due to sniffer tip
► To prevent injury from accidental contact with the sniffer tip, align the sniffer tip in
the holder so that it points away from the operator.
A bracket is available for the sniffer tip. The holder can be mounted on the front of the
device, see also “Device setup [}14]“.