INFICON Installation | 5
XL3000flex-Operating-instructions-jina83en1-07-(2105) 31 / 108
ü You have a separately available I/O module or bus module. See also "Accessories
[}101]".
ü You have a commercially available DIN-TS35 DIN rail.
1
Place the sniffer leak detector at least 20 cm away on all sides.
2
Connect the SL3000XL sniffer cable to the front of the XL3000flex.
3
Fasten the DIN rail using the screws already located in the threaded holes on
the back of the device (see Device setup [}14]).
4
Then hook the module into the DIN rail at the bottom, press it against the DIN
rail at the top and let it snap into place.
5
Connect the I/O module or the bus module via a data cable to the LD socket on
the back of the XL3000flex.
See also
2 Selecting the type of expansion module [}45]
2 Assigning analog outputs of the I/O module [}46]
2 Settings for bus module BM1000 [}53]
5.4 Fasten holder for sniffer line (optional)
WARNING
Danger to pacemaker wearers from magnets
The function of a pacemaker can be influenced by the magnet on the back of the
holder.
If you wear a pacemaker, do not install it yourself.
If you wear a pacemaker, always maintain a distance of at least 10 cm from the
holder when operating the device.
CAUTION
Risk of injury due to sniffer tip
To prevent injury from accidental contact with the sniffer tip, align the sniffer tip in
the holder so that it points away from the operator.
A bracket is available for the sniffer tip. The holder can be mounted on the front of the
device, see also “Device setup [}14]“.