6 | Operation INFICON
36 / 108 XL3000flex-Operating-instructions-jina83en1-07-(2105)
5 g/a
6 oz/yr
7 sccm
8 sft
3
/yr
Control unit Settings > Set up > Interfaces > Units (interface) > Leak rate unit
SNIF
LD protocol Command 432 (sniffing)
ASCII protocol Command *CONFig:UNIT:LRSnif
6.2.4 Select display unit for pressure
Device of pressure
mbar
Pa
atm
Torr
Control unit Main menu > Display > Units (display) > Pressure unit
6.2.5 Select interface unit for pressure
Pressure unit
interface
Selecting the pressure device of the interfaces
0 mbar (factory setting)
1 Pa
2 atm
3 Torr
Control unit Main Menu > Settings > Setup > Interfaces > Units (Interface) >
Pressure Unit
LD protocol Command 430 (sniffing)
ASCII protocol Command *CONFig:UNIT:Pressure
6.2.6 Settings for the XL sniffer adapter
For operation with the XL Sniffer Adapter you have to use the
– SL3000XL sniffer line,
– Select the "XL Sniffer Adapter" operation mode, see "Operation mode
“Sniffing” [}53]".