INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 53 / 108
Function State: Description
Measure closed:
open:
Measuring (ZERO is possible, all trigger outputs switch depending on the leak
rate.)
Standby or emission disabled (ZERO is not possible, all trigger outputs will
return "Leak rate threshold value exceeded".)
Standby closed:
open:
Standby (ZERO is not possible, all trigger outputs will return “Leak rate
threshold value exceeded”.)
Measuring (ZERO is possible, all trigger outputs switch depending on the leak
rate.)
SNIF closed:
open:
SNIF
VAC
Error or
warning
closed:
open:
Error or warning
No error or warning
Gas ballast closed:
open:
Gas ballast is active
Gas ballast is inactive
Calibration
leak open
closed:
open:
calibration leak is active
calibration leak is inactive
CAL stable closed:
open:
Calibration completed with calibration leak (see "Time and general
preferences [}60]")
Assignment not stable or calibration is inactive
Cathode 2 closed:
open:
Cathode 2 is active
Cathode 1 is active
6.2.24 Settings for bus module BM1000
Address of bus module
Setting the bus module address. (Node address with Profibus, MACID with
DeviceNet)
0 ... 255
Control unit Settings > Set up > Interfaces > Bus module > Address
LD protocol 326
ASCII protocol –
6.2.25 Operation mode “Sniffing”
The device has the sniffer mode with a high flow rate.
For the XL3000flex , only this operation mode makes sense.