6 | Operation INFICON
52 / 108 XL3000flex-Operating-instructions-jina83en1-07-(2105)
LD protocol Command 263
ASCII protocol *CONFig:PLCOUTLINK:1 (2 ... 8)
Functions, assignment of digital outputs:
Function State: Description
Open open: always open
Trigger 1 closed:
open:
Value exceeded leak rate threshold Trigger 1
Value fell below leak rate threshold Trigger 1
Trigger 2 closed:
open:
Value exceeded leak rate threshold Trigger 2
Value fell below leak rate threshold Trigger 2
Trigger 3 closed:
open:
Value exceeded leak rate threshold Trigger 3
Value fell below leak rate threshold Trigger 3
Trigger 4 closed:
open:
Value exceeded leak rate threshold Trigger 4
Value fell below leak rate threshold Trigger 4
Ready closed:
open:
Emission switched on, calibration process inactive, no error
Emission switched off or calibration process active or error
Warning closed:
open:
Warning
no warning
Error closed:
open:
Error
no error
CAL active closed:
open:
Device is to be calibrated.
Device is not to be calibrated.
CAL request closed:
closed:
open:
and no external calibration: Calibration request (with temperature change from
5°C or 30 minutes after the start-up or if default rotation speed was changed)
and external calibration or "CAL check": Request "Open or close external
calibration leak"
no request
Run-up closed:
open:
Run-up
no run-up
ZERO active closed:
open:
ZERO switched on
ZERO switched off
Emission on closed:
open:
Emission switched on
Emission switched off