INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 51 / 108
Function Flank/state: Description
Internal PROOF inactive→
active:
Start the internal Proof function. In the XL3000flex without function.
External
PROOF
inactive→
active:
Start the external Proof function.
START / STOP
impulse
inactive→
active:
Activate Start or Stop.
ZERO update inactive→
active:
active→
inactive:
Update or switch on ZERO
No function
Flow inactive→
active:
active→
inactive:
Switch flow of SL3000XL to 3000 sccm (XL adapter)
Switch flow of SL3000XL to 300 sccm (XL adapter)
CAL machine inactive→
active:
Determining the machine factor or of the sniff factor
External CAL
check
inactive→
active:
Check calibration with external calibration leak
Start / Stop
impulse
inactive→
active:
Switching between measuring operation and standby
Mass 2 / Mass
4
inactive→
active:
active→
inactive:
Activate mass 4
Activate mass 2
Photo
interrupter
inactive→
active:
active→
inactive:
Sniffer tip at calibration leak, calibration starts.
Sniffer tip has been removed.
6.2.23 Assigning the digital outputs of the I/O module
The digital outputs PLC-OUT 1 ... The available functions can be assigned in any
way necessary to the 8 I/O module.
Every function can be inverted.
Possible functions: see the following table
Control unit Settings > Set up > Interfaces > I/O module > Digital outputs >
Configuration PLC Output