Control unit Settings > Set up > Operation modes > Sniff > Capillary > Blocked
XL > Pressure capillary blocked XL
LD protocol Command 455
ASCII protocol Command *CONFig:PRESSXLLow
Pressure value XL
capillary broken (high
flow)
You set a maximum pressure value in order to detect a disruption in the XL capillary
(high flow, 3000 sccm). If the value is exceeded, the system issues warning 552.
200 ... 600 mbar
Control unit Settings > Set up > Operation modes > Sniff > Capillary > Broken
XL > Pressure capillary broken XL
LD protocol Command 456
ASCII protocol Command *CONFig:PRESSXLHigh
Select flow
Select low flow or high flow. Comment: The selection can also be made with the right
sniffer key or assigned to one of the favorite keys of the control unit.
Small (low flow)
Large (high flow)
Control unit Settings > Configuration > Operating Mode > Flow > Flow Control
or Functions > Flow > Flow Control
LD protocol Command 229
ASCII protocol Command *CONFig:Highflow
6.2.7 Operator types and authorizations
There are four different operator types that are distinguished by different
authorizations. The integrator is registered ex works.
Additional operators can be registered. The following table shows options for
individual operator types to register new operator types.
Operator registration
Viewer Operator Supervisor Integrator
- Operator
Viewer
Supervisor
Operator
Viewer
Integrator
Supervisor
Operator
Viewer
For the types "Integrator", "Supervisor" and "Operator", a four-digit PIN must be
assigned during registration (0000 ... 9999). "0000" is assigned to all operators ex
works.