INFICON Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105) 63 / 108
Fig.5:
External calibration with IO1000 using the example of sniffer line SL3000XL, description
of PLC inputs and outputs: "Assigning analog outputs of the I/O module [}46]"
6.3.4.3 Check the calibration
To check whether a re-calibration is necessary, check the already existing.
6.3.4.4 Calibration using the external calibration leak test
► LD and ASCII protocol: The status must be queried via: Command 260 or
*STATus:CAL
1
Hold the sniffer line to the test leak.
2
Wait until leak rate signal is tuned and stable.