9–42 Minimate Pro Operator Manual
Service Technician can assist you in determining how to repair or describe a
workaround for returning the monitor to a working state.
(1) Press the Down Arrow key to scroll and highlight the Self Check menu option.
(2) Press the Right Arrow button to activate Self Check – Now.
(3) Press the Enter key to Continue to each screen until you are returned to the
main display. Review each screen to ensure a Fail does not appear.
(4) A list of diagnostic results will appear with a Pass or Fail shown next to them.
Only items listed with a Fail result will appear in the unit Monitor Log.
f. Disconnect Battery
Use the Disconnect Battery feature if you are planning to store the unit, prepare it for
transportation for service or calibration, or if the unit is not performing to standards.
(1) Select the Setup key on the unit.
(2) Press the Down Arrow to scroll and highlight the Maintenance Menu. Press
the Enter key to accept the selection.
(3) A message will appear, “Disconnecting the battery will shut the unit down.
Press the Right Arrow key to proceed, or any other key to cancel.” Press the
Right Arrow key and the unit will power off.
g. Zero Offsets
The Zero Offsets menu will monitor the geophone and microphone for 10 seconds and
then analyze the data to determine if there is an offset on any of the channels. If an offset
has been detected, the value associated with this offset will be stored in the unit. This
offset will then be used to correct the data being recorded by adding the stored offset
value to the data. This corrected value is then stored as the event data.
Caution: Ensure there are no vibration or sound signals produced while running this
function as it will negatively affect the results.
(1) Select Zero Offsets from the Maintenance Menu. The following menu appears: