944 Minimate Pro Operator Manual
2) If the measured signal is greater than 94 dB: “Error Calibrating Sound
Level Microphone. Calibration reference signal too large. Existing
Calibration retained.”
Press the Enter key to continue.
(11) If the monitor and field calibration instrument are unable to calibrate the
microphone, the following message appears, “Error calibrating the Sound
Level Microphone. Please send the Microphone to be serviced.
(12) If the field calibration was successful, the following message appears,
Calibration successful. Do you wish to save the new calibration results?”
(13) Press the Right Arrow to save the calibration results, or press Cancel to retain
the original calibration date. If the changes are accepted, the information will
appear on the Event Report showing the Company Name of the Calibrator in
the Field Calibration” field. Refer to the Blastware Operator manual for more
information.