TL6000, Users
Rev L, June 2008
Page 19 of 22
Error Messages
OvER
OvEROvER
OvER
The scale is over capacity. Reduce the weight applied to the scale.
OE
OEOE
OE
The scale is outside the internal A/D converter range. Reduce the weight
applied to the scale.
dISP
dISPdISP
dISP
Display error. Value is too large to fit on the display. Reduce the weight
applied to the scale, or try changing graduations.
-dISP
-dISP-dISP
-dISP
Display error. Value is too negative to fit on the display. Press ZERO to
zero the scale.
EEPE
EEPEEEPE
EEPE
EEPROM error. The scale has had it’s calibration corrupted or destroyed;
the scale will require calibration.
Err 1
Err 1Err 1
Err 1
Error during calibration. The scale automatically exits calibration mode.
When calibrating, make sure that each of the three calibration weights is
larger than the previous weight.
Err 2
Err 2Err 2
Err 2
EEPROM error. The scale is unable to read the EEPROM.
Err 4
Err 4Err 4
Err 4
EEPROM error. The scale is unable to write to the EEPROM.
CALSt
CALStCALSt
CALSt
This message will show for a few seconds if you have attempted to access
weight calibration with the calibration strap in place. See the “Calibration”
section if you need to calibrate.
Troubleshooting
Problem: Solution:
Display freezes or is blank and no
response to keys pressed
The MCU may have locked up. Try
removing both batteries together, and then
replacing one or both of them. As a
general rule to reduce lock-ups, whenever
turning the TL6000 off, wait at least 5-10
seconds before turning it on again.
Batteries are good but the scale won’t
power up.
If the scale is turned off and turned back
on right away, the scale may not be able to
power up. Wait about 10 seconds after
powering off before pressing the ON key
again.