TL6000, Users
Rev L, June 2008
Page 9 of 22
Note: If your scale has the optional audio alarm, Set Point #1 and #2 are
not wired to the external connector. Instead, Set point #1 controls when
the alarm will sound. The alarm will sound whenever the weight applied to
the scale is greater than or equal to the weight entered into “Set P1”.
Changing Set Points
Press the MODE button three times until the display reads ‘SEtP1
SEtP1SEtP1
SEtP1’. Press
and hold the mode button until it reads ‘Edit
EditEdit
Edit’. Change the value by using
the up/down, left/right (MODE/TARE, PEAK/ZERO) buttons. To save the
set point press and hold the MODE button until it reads ‘SAvE
SAvESAvE
SAvE’. The
TL6000 will return to measurement mode. To change Set Point 2, press
the MODE button four times, until the display reads ‘SEtP2
SEtP2SEtP2
SEtP2’ and follow the
same instructions as above instructions.
Default = 99999.
SAMPLE RATE
The higher the sample rate, the more averaging of past readings the
tension link uses to compute the displayed weight. ‘1’ results in the
fastest update times, while ‘100’ gives you the most stable read-outs.
Changing the Sample rate:
Press the MODE key five times until the display reads ‘S rt
S rtS rt
S rt ’. Press and
hold the MODE key until the display reads ‘Edit
EditEdit
Edit’. Change the value by
using the arrow keys. To save the new sample rate press and hold
MODE until the display reads ‘Save
SaveSave
Save’. The TL6000 will return to
measurement mode.
Default = 5.
SLEEP
The TL6000 features a low-power mode to conserve battery power. If the
scale does not experience any activity for a predetermined amount of
minutes, the system will switch to low -power mode. When in low power
mode, the scale still operates normally. The only difference is a slowed
response rate. You may have to wait a second or two for a change in
weight to show up on the display. Any key pressed or change in weight
will return the scale to normal mode.
Adjusting Sleep:
Press the MODE key six times until the display reads ‘SLEEP
SLEEPSLEEP
SLEEP’. Press and
hold the MODE key until the display reads ‘Edit
EditEdit
Edit’. Change the value (in
minutes) using the arrow keys. Any entry from 1 to 180 minutes is
allowed. To disable the sleep feature enter 0. To save your data press
and hold the MODE key until the display reads ‘Save
SaveSave
Save’. The system will
return to measurement mode.
Default = 5.