3 EMC INFORMATION
Invacare ®CS Series Beds
18 Part No 1153410
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A / m 3 A / m Power frequency magnetic fields should be at levels characteristic of
a typical location in a typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The CS Bed is intended for use in the electromagnetic environment specified below. The customer or the user of the CS Bed should assure that it is used
in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz 3 Vrms
Portable and mobile RF communications equipment should be used no closer to
any part of the CS Bed, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d=1.2
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
P