Invacare®EtudePlus
qCheckthefunctionofthebrakes.
qCheckthelockingfunction(s)ofthesiderails.
qCheckthatallbedslatsareintactandnotloose.
6.5CleaningandDisinfection
6.5.1GeneralSafetyInformation
CAUTION!
RiskofContamination
–Takeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
RiskofElectricShockandProductDamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
NOTICE!
Wronguidsormethodscanharmordamage
theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
–Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend
anordinaryhouseholdcleaningagentsuchas
dishwashingliquid,ifnotspeciedotherwisein
thecleaninginstructions.
–Neveruseasolvent(cellulosethinner,acetone
etc.)thatchangesthestructureoftheplastic
ordissolvestheattachedlabels.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
6.5.2CleaningIntervals
NOTICE!
Regularcleaninganddisinfectionenhancesmooth
operation,increasestheservicelifeandprevents
contamination.
Cleananddisinfecttheproduct:
–regularlywhileinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser.
6.5.3CleaningbyHand
Allcomponents(excludingremovabletextiles)
•Cleaningagent:Werecommendusingamilddetergent
withneutralornearneutralpH(5-9).
Commerciallyavailableconsumerproductslike
handdishwashingliquidorall-purposecleaners
canbeused.Readtheinstructionsonthe
labelofyourcleaningagentanduseitatthe
indicatedconcentration.
•Max.watertemperature:40°C
1.Wipesurfacesthoroughlywithasoftclothslightly
moistenedwithdetergentsolutionuntilallvisiblesoil
isremoved.
2.Removealldetergentresiduewithacleandampcloth.
3.Wipedryusingacleansoftcloth.
Removabletextiles(includingupholsteryandmattresses)
•Seeattachedlabeloneachproduct.
6.5.4DisinfectionInstructions
InDomesticCare
•Disinfectant:Werecommendusinganalcohol-based
surfacedisinfectant(with70-90%alcohol).
Readtheinstructionsonthelabelofyour
disinfectant.Itprovidesinformationonthe
activityspectrum(bacteria,fungiand/orviruses),
materialcompatibilityandthecorrectexposure
time.
1.Ensurethesurfacesarecleanedbeforedisinfection.
2.Moistenasoftclothandwipe-disinfectallaccessible
surfacesandkeepthemmoistenedfortheexposure
timestatedonthelabelofthedisinfectant.
3.Allowtheproducttoair-dry.
InInstitutionalCare
Followyourin-housedisinfectionproceduresandonlyuse
thedisinfectantsandmethodsspeciedtherein.
6.6Lubrication
Werecommendlubricatingthebedaccordingtothe
followinginstructionafterserviceandcleaningorwhenever
itisnecessarytoreducefriction.
PartofbedLubricationmethod
Pointsofrotationinmattress
support
Oil(medicallyclean)
Motorattachmentpointson
mattresssupport
Oil(medicallyclean)
Motorbearingsandtension
rods
Oil(medicallyclean)
Thewoodensiderailsglidingsystemmustnotbe
lubricatedwithoil-otherwisethewoodenbarswill
movesluggishly.
Fororderingofcorrectoil,contactyourInvacare
provider .
2060128467-B