EasyManua.ls Logo

Invacare Etude Plus - Cleaning, Disinfection, and Lubrication; Cleaning and Disinfection Procedures

Invacare Etude Plus
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
InvacarEtudePlus
qCheckthefunctionofthebrakes.
qCheckthelockingfunction(s)ofthesiderails.
qCheckthatallbedslatsareintactandnotloose.
6.5CleaningandDisinfection
6.5.1GeneralSafetyInformation
CAUTION!
RiskofContamination
Takeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
RiskofElectricShockandProductDamage
Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
Makesurethatnowatersplashestotheplug
orthewalloutlet.
Donottouchthepowersocketwithwethands.
NOTICE!
Wronguidsormethodscanharmordamage
theproduct.
Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend
anordinaryhouseholdcleaningagentsuchas
dishwashingliquid,ifnotspeciedotherwisein
thecleaninginstructions.
Neveruseasolvent(cellulosethinner,acetone
etc.)thatchangesthestructureoftheplastic
ordissolvestheattachedlabels.
Alwaysmakesurethattheproductiscompletely
driedbeforetakingintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
6.5.2CleaningIntervals
NOTICE!
Regularcleaninganddisinfectionenhancesmooth
operation,increasestheservicelifeandprevents
contamination.
Cleananddisinfecttheproduct:
regularlywhileinuse,
beforeandafteranyserviceprocedure,
whenithasbeenincontactwithanybody
uids,
beforeusingitforanewuser.
6.5.3CleaningbyHand
Allcomponents(excludingremovabletextiles)
Cleaningagent:Werecommendusingamilddetergent
withneutralornearneutralpH(5-9).
Commerciallyavailableconsumerproductslike
handdishwashingliquidorall-purposecleaners
canbeused.Readtheinstructionsonthe
labelofyourcleaningagentanduseitatthe
indicatedconcentration.
Max.watertemperature:40°C
1.Wipesurfacesthoroughlywithasoftclothslightly
moistenedwithdetergentsolutionuntilallvisiblesoil
isremoved.
2.Removealldetergentresiduewithacleandampcloth.
3.Wipedryusingacleansoftcloth.
Removabletextiles(includingupholsteryandmattresses)
Seeattachedlabeloneachproduct.
6.5.4DisinfectionInstructions
InDomesticCare
Disinfectant:Werecommendusinganalcohol-based
surfacedisinfectant(with70-90%alcohol).
Readtheinstructionsonthelabelofyour
disinfectant.Itprovidesinformationonthe
activityspectrum(bacteria,fungiand/orviruses),
materialcompatibilityandthecorrectexposure
time.
1.Ensurethesurfacesarecleanedbeforedisinfection.
2.Moistenasoftclothandwipe-disinfectallaccessible
surfacesandkeepthemmoistenedfortheexposure
timestatedonthelabelofthedisinfectant.
3.Allowtheproducttoair-dry.
InInstitutionalCare
Followyourin-housedisinfectionproceduresandonlyuse
thedisinfectantsandmethodsspeciedtherein.
6.6Lubrication
Werecommendlubricatingthebedaccordingtothe
followinginstructionafterserviceandcleaningorwhenever
itisnecessarytoreducefriction.
PartofbedLubricationmethod
Pointsofrotationinmattress
support
Oil(medicallyclean)
Motorattachmentpointson
mattresssupport
Oil(medicallyclean)
Motorbearingsandtension
rods
Oil(medicallyclean)
Thewoodensiderailsglidingsystemmustnotbe
lubricatedwithoil-otherwisethewoodenbarswill
movesluggishly.
Fororderingofcorrectoil,contactyourInvacare
provider .
2060128467-B

Table of Contents