EasyManua.ls Logo

Invacare Etude Plus - RF Wireless Communications Immunity

Invacare Etude Plus
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
InvacarEtudePlus
Immunitytest
Test/Compliancelevel
Electromagneticenvironmentguidance
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
<0%U
T
for0,5cycleat
45°steps
0%U
T
for1cycles
70%U
T
for25/30cycles
<5%U
T
for250/300
cycles
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.Iftheuserofthisproductrequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthattheproductispowered
fromanun-interruptiblepowersupplyorabattery.
U
T
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
30A/m
Powerfrequencymagneticeldsshouldbeatlevelscharacteristicofa
typicallocationinatypicalcommercialorhospitalenvironment.
ConductedRF
IEC61000-4-6
3V
150kHzto80Mhz
6V
inISM&amateurradio
bands
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio
(cellular/cordless)telephonesandlandmobileradios,amateurradio,AM
andFMradiobroadcastandTVbroadcastcannotbepredictedtheoretically
withaccuracy.ToassesstheelectromagneticenvironmentduetoxedRF
transmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe
measuredeldstrengthinthelocationinwhichthisproductisusedexceeds
theapplicableRFcompliancelevelabove,thisproductshouldbeobserved
toverifynormaloperation.Ifabnormalperformanceisobserved,additional
measuresmaybenecessary,suchasreorientingorrelocatingthisproduct.
Interferencemayoccurinthevicinityofequipmentmarkedwiththe
followingsymbol:
RadiatedRF
IEC61000-4-3
10V/m
80Mhzto2,7GHz
385MHz-5785MHz
testspecications
forimmunitytoRF
wirelesscommunication
equipmentrefertotable
9ofIEC60601-1-2:2014
PortableandmobileRFcommunicationsequipmentshouldbeusedno
closerthan30cmtoanypartofthisproductincludingcables.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection
fromstructures,objectsandpeople.
10.3.1TestspecicationsforimmunitytoRFwirelesscommunicationsequipment
IEC60601-1-2:2014Table9
TestFrequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power(W)
Distance(m)
Immunitytest
level(V/m)
385380-390TETRA400
Pulse
modulation
b)
18Hz
1.80.327
450430-470
GMRS460,
FRS460
FM
c)
±5kHz
deviation
1kHzsine
20.328
710
745
788
704-787
LTEBand13,17
Pulse
modulation
b)
217Hz
0.20.39
810
870
930
800-960
GSM800/900,
TETRA800,
iDEN820,
CDMA850,LTE
Band5
Pulse
modulation
b)
18Hz
20.328
2660128467-B

Table of Contents