Electromagneticcompatibility(EMC)
10Electromagneticcompatibility(EMC)
10.1GeneralEMCinformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccordingtotheEMCinformationinthismanual.
ThisproducthasbeentestedandfoundtocomplywithEMClimitsspeciedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffecttheoperationofthisproduct.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabove
standard.Todetermineiftheemissionfromthisproductiscausingtheinterference,runandstoprunningthisproduct.If
theinterferencewiththeotherdeviceoperationstops,thenthisproductiscausingtheinterference.Insuchrarecases,
interferencemaybereducedorcorrectedbythefollowing:
•Reposition,relocate,orincreasetheseparationbetweenthedevices.
10.2Electromagneticemission
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11
GroupI
ThisproductusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11
ClassB
Thisproductissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
10.3ElectromagneticImmunity
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Immunitytest
Test/Compliancelevel
Electromagneticenvironment–guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±8kVcontact
±2kV,±4kV,±8kV,±
15kVair
Floorsshouldbewood,concreteorceramictile.Ifoorsarecoveredwith
syntheticmaterial,therelativehumidityshouldbeatleast30%.
Electrostatic
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines;100kHzrepetition
frequency
±1kVforinput/output
lines;100kHzrepetition
frequency
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Surge
IEC61000-4-5
±1kVlinetoline
±2kVlinetoearth
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
60128467-B25