InvacarEtudePlus
9.5Environmentalconditions
Storageand
Transportation
Operation
Temperature
-10°Cto+50°C+5°Cto+40°C
Relativehumidity
20%to80%-notcondensing
Atmospheric
pressure
700hPato1060hPa
Allowtheproducttoreachoperationtemperature
beforeusage:
Warmingupfromminimumstoragetemperature
mighttakeupto24hours.
Coolingdownfrommaximumstorage
temperaturemighttakeupto24hours.
9.6Electricalsystem
Voltage
supply
Uin=230Volt,AC,50/60Hz(AC=
Alternatingcurrent)
Maximum
currentinput
Iinmax.=1.5A
Intermittent
(periodic
motor
operation)
Int=10%,max.2min/18min
Degreeof
protection
IPX6 or IPX4 (depending on conguration)
When the bed is equipped with electrical
leg rest it is protected according to IPX4.
IPX6 The electrical system is protected
against water projected in powerful jets
from any direction.
IPX4 The electrical system is protected
against water splashed against the
components from any direction.
Insulation
class
ClassIIequipment
TypeB
AppliedPart
AppliedPartcomplyingwiththespecied
requirementsforprotectionagainstelectrical
shockaccordingtoIEC60601-1.
Soundlevel
45to50dB(A)
Thebedhasnoisolator(mainswitch).Ifthebed
needstobeelectricallydisconnected,ithastobe
unpluggedfromthemainssocket.
24
60128467-B

Table of Contents