Invacare®EtudePlus
9.5Environmentalconditions
Storageand
Transportation
Operation
Temperature
-10°Cto+50°C+5°Cto+40°C
Relativehumidity
20%to80%-notcondensing
Atmospheric
pressure
700hPato1060hPa
Allowtheproducttoreachoperationtemperature
beforeusage:
•Warmingupfromminimumstoragetemperature
mighttakeupto24hours.
•Coolingdownfrommaximumstorage
temperaturemighttakeupto24hours.
9.6Electricalsystem
Voltage
supply
Uin=230Volt,AC,50/60Hz(AC=
Alternatingcurrent)
Maximum
currentinput
Iinmax.=1.5A
Intermittent
(periodic
motor
operation)
Int=10%,max.2min/18min
Degreeof
protection
IPX6 or IPX4 (depending on conguration)
When the bed is equipped with electrical
leg rest it is protected according to IPX4.
IPX6 – The electrical system is protected
against water projected in powerful jets
from any direction.
IPX4 – The electrical system is protected
against water splashed against the
components from any direction.
Insulation
class
ClassIIequipment
TypeB
AppliedPart
AppliedPartcomplyingwiththespecied
requirementsforprotectionagainstelectrical
shockaccordingtoIEC60601-1.
Soundlevel
45to50dB(A)
Thebedhasnoisolator(mainswitch).Ifthebed
needstobeelectricallydisconnected,ithastobe
unpluggedfromthemainssocket.
24
60128467-B