8 SAFETY INSPECTION/MAINTENANCE
Part No 1154295 75 Invacare® Solara®3G/Spree 3G Wheelchair
Inspect tilt slides and roller bearings to ensure they are free from dirt, lint, etc.
Inspect seat positioning strap for any signs of wear. Ensure buckle latches. Verify hardware that attaches strap to frame is secure and undamaged.
Replace if necessary.
Inspect/Adjust Periodically
Inspect clothing guards for bent or protruding metal.
Ensure that all fasteners on clothing guards are secure.
Ensure arms are secure but easy to release and adjustment levers engage properly.
Ensure that casters are free of debris.
Ensure adjustable height arms operate and lock securely.
Ensure armrest upholstery has no rips.
Ensure armrest pad sits flush against arm tube.
Ensure trigger release lever cables and handles return when released.
Ensure sealed bearings and axle nut tension are correct.
Inspect handrims for signs of rough edges or peeling finish.
Ensure seat, back and/or arm upholstery have no rips.
Clean upholstery and armrests.
Clean and wax all parts.
Check that all labels are present and legible. Replace if necessary.
CAUTION
Damage from Bleach
Clean upholstery with mild soap and water or spray disinfectant using a sponge. DO NOT use bleach or wash in a washing
machine.