Service
6Service
6.1Tighteningtorques
CAUTION!
Damagecanbecausedtothemobilitydevice
duetoimproperlytightenedscrews,nutsor
plasticconnections.
–Alwaystightenscrews,nutsetctothestated
tighteningtorque.
–Onlytightenscrewsornutswhicharenotlisted
herengertight.
Thetighteningtorquesstatedinthefollowinglistarebased
onthethreaddiameterforthenutsandboltsforwhichno
specicvalueshavebeendetermined.Allvaluesassume
dryandde-greasedthreads.
ThreadTighteningtorquein
Nm±10%
M43Nm
M5
6Nm
M610Nm
M825Nm
M1049Nm
M1280Nm
M14120Nm
M16180Nm
UNCThreadTighteningtorquein
Nm±10%
1/4”-20
11–7Nm
5/16”-20
22–14Nm
3/8”-16
41–25Nm
7/16”-14
67–40Nm
1/2”-13
100–60Nm
9/16”-12
150–90Nm
5/8”-11
210–130Nm
3/4”-1
370–230Nm
7/6”-9
600–370Nm
1”-8
900–550Nm
6.2Overviewmobilitydevice
ThisoverviewappliesforTDXSP2mobilitydeviceswith
ModuliteSeat.
Underneathseat
BatteriesbehindfrontshroudA
Fig.6-1
PowermodulebehindrearshroudA
Fig.6-2
Actuatormodules
Tiltactuator(optional)A
Actuatormodule(optional)B
Lifteractuator(optional)C
Fig.6-3
Lifter
ACTwithLifter
1532486-L13