Invacare®TDX®SP2Series
Symptom
PossiblecauseSolution
BatteriesdonotchargeFuseonmobilitydevicehasburntoutCheckmobilitydevicefuses.See6.7.13
Checkingandreplacingmainfuse,page
37.
Chargernotconnectedtomobility
device
Makesurethatchargerhasbeen
connectedtomobilitydevice
NomainssupplyCheckmainssupplywithvoltmeter .
DefectivemainssupplycableCheckmainssupplycable.Replace
damagedcablesorsendchargerto
Invacare®Serviceforrepair
BatterychargercouldbedefectiveUsebatterychargerwhichyouknowis
workingproperlytochargebatteries.
Senddefectivebatterychargerto
Invacare®Serviceforrepair.
Batteryvoltageistoolowtooperate
mobilitydevice.
Replacebatteries.See6.7.9Replacing
batteries/replacingbatterycables,page
35.
7.4Serviceplan(1xannually)
CAUTION!
Riskofinjuryanddamagetoproperty,ifthemaximumspeedreductiononawheelchairwithalifterdoesnot
functioncorrectly
Thewheelchair’scontrolunitmustreducethemaximumpossiblespeedassoonasthelifterisraised.
–Testthemaximumspeedreductionforcorrectfunctionafteranymaintenanceworkormodicationstothe
wheelchair.
Component
CheckRemedy
Notes
DamagetoposturebeltReplacebeltif
damaged
See“Replacingposturebelt”in
ModuliteServiceManual.
Posturebelt
BeltlockfunctionReplacebeltif
damaged
See“Replacingposturebelt”in
ModuliteServiceManual.
Damagetoarmrests
Replacecoveringif
damaged
Armrests
ArmrestxingsTightenscrews
DamagetomudguardReplacemudguard,if
damaged
Mudguard
SidepanelxingsTightenscrews
PoweredtiltTighttofSLretainer
clip
ReplaceSLretainer
clip,ifnecessary
Poweredrecline(if
tted)
Damagetobackrest
Seams
Fixing
Checkcable
Checkfunction
Replacepartsif
damaged
Tightenscrews
Replacecableormotor
ifnecessary
Frames(chassis)/
batterymounting
Checkxings,welded
seamsandbattery
mounting
Tightenscrews
Replacecomponentsif
necessary
StabilityLockCheckgascylinderRepairorreplaceif
damaged
Replacegascylinder
every2years!
See6.3.1Stabilitylock,page14.
461532486-L