Page #50 background imageLoading...
Page #50 background image
50
Problem
Possible cause
Solution
Display remains blank
Battery flat
Check the connection to the power
supply and make certain the switch is
in position 'I'
Adapter damaged
Contact Inventis service department or
dealer
Battery does not
recharge
USB cable damaged
Change the USB cable (USB A mini
B standard)
Adapter damaged
Contact Inventis service department or
dealer
Instrument not
positioned correctly in
charging base
Check the positioning of the instrument
Check that the contacts are clean
Battery damaged
Replace the battery - Contact Inventis
service department or dealer
A test cannot be
accessed
Optional test not
activated
Contact the Inventis technical service
department to obtain the license,
supplying the serial number of the
instrument
message:
“Hardware error”
Non-fatal internal error
Press OK to continue; if the problem
persists, contact the Inventis service
department
message:
“Serious error”
Fatal internal error
Restart the instrument; if the problem
persists, contact the Inventis service
department