Appendix C
Electromagnetic compatibility
The instrument has been tested and been found to comply with the limits for
electromedical devices imposed by IEC 60601-1-2 standard. These limits
guarantee a reasonable level of protection against harmful interference in a
typical medical installation.
The instrument generates, utilizes and radiates energy at radio frequency,
and if not correctly installed and used as indicated in the instructions, can
cause harmful interference to other devices located in the immediate
vicinity. At all events, there can be no guarantee that interference will not
occur in certain situations.
This instrument is suitable for use in professional healthcare facilities, for
example in hospital buildings, although not in close proximity to high
frequency surgical equipment and RF shielded rooms housing MRI systems,
where the intensity of electromagnetic disturbance is typically high.
The instrument must not be used in close proximity to other
equipment or stacked together with other equipment. If this cannot
be avoided, check the operation of the instrument carefully to
ensure that it is able to deliver normal levels of performance in the
configuration adopted.
Should the instrument appear to interfere with other devices — this can be
verified by switching off the instrument and then switching on again — it
may be possible to remedy the problem by adopting one or more of the
following solutions:
- change the orientation and/or the position of the device affected by
interference;
- distance the devices one from another;
- consult the manufacturer or a service centre for assistance.
Cables, transducers and accessories
The cables, transducers and accessories for which Inventis declares
compliance with IEC 60601-1-2 standard are those supplied with the
device, in particular:
1) Medical grade USB power adapter
2) Shielded USB cable, maximum length: 2 m
3) Charging base